DTS140 - a universal system for display measurement and more

Complete systems based on the CAS 140CT array spectrometer featuring a high-end, back-illuminated CCD detector are available for measurements of radiance and luminance as well as color of displays and back-illuminated panel graphics. No matter what kind of display technology has to be tested - whether LCD, CRT, OLED or LED displays - the DTS140 can be used universally.  


Coupling of light into the system is achieved using the TOP 200 telescopic optical probe that is attached to the spectrometer via an optical fiber. A large selection of lenses is available for the TOP 200 in order to provide different measuring spot sizes. The new SpecWin Pro software offers comprehensive tools for data analysis and documentation of measurement results.

The fiber coupling of the TOP 200 makes it possible to simply swap the adapter and quickly reconfigure the DTS140 for other applications (e.g. LED measurement). The range of applications available is thus considerably expanded - a feature hard to realize for spectrometers with a fixed, integrated lens.

Features:

Specifications

DTS140-131

DTS140-133

DTS140-231

DTS140-233

Spectral range

360 - 830 nm

380 - 1040 nm

360 - 830 nm

380 - 1040 nm

Spectral resolution

2.2 nm

3.0 nm

2.2 nm

3.0 nm

Datapoint interval

0.5 nm

0.65 nm

0.5 nm

0.65 nm

TOP 200 Telescopic Optical Probe

Lens

60 mm

60 mm 

HRL 90

HRL 90

Working distance *)

18 cm 

18 cm

23 cm

23 cm

Stray light (approx.)

1 %

1 %

0.1 %

0.1 %

Measurement

sensitivity range

[cd/m²]

Aperture 1

0.2 - 1*109

0.1 - 1*109

0.6 - 4*109

0.4 - 3*109

Aperture 2

0.06 - 4*108

0.04 - 3*108

0.2 - 1*109

0.1 - 1*109

Aperture 3

0.02 - 1*108

0.01 - 1*108

0.06 - 4*108

0.04 - 3*108

Aperture 4

0.006 - 4*107

0.004 - 3*107

0.02 - 1*108

0.01 - 1*108

Accuracy

Luminance

± 3 %

Radiance

± 4 %

Color coordinates (x,y)

± 0.0015

Dominant wavelength

± 0.5 nm

Polarization sensitivity

< 1 %

*) Distance from the test sample to the TOP 200's front panel; for models with the 60 mm lens, basic calibration is done at a working distance of 50 cm.